<Range low> = MA_LOW_XX = 0.004 [A] (default value)
<Range high> = MA_HIGH_XX = 0.020 [A] (default value)
<Temperature low> = MA_TEMP_LOW_XX = 0 [°C] (default value)
<Temperature high> = MA_TEMP_HIGH_XX = 100 [°C] (default
value)
CH_CONF MV (@<ch_list>)
Sets the selected for measuring mV.
CH_CONF <TC type>, <CJC type>, <Manuel CJC temperature>
<temp. unit>, <A>, <B>, <C> (@<ch_list>)
Sets the selected channels for measuring thermocouple
<TC type> can be :
J, K, T, E, R, S, B, XK, BP, L, U, N,
CUSTOM J,
CUSTOM K,
CUSTOM T,
CUSTOM E,
CUSTOM R,
CUSTOM S,
CUSTOM B,
CUSTOM XK,
CUSTOM BP,
CUSTOM L,
CUSTOM U,
CUSTOM N
<CJC type> can be : INT, EXT, MAN
<Manuel CJC temperature> is only used with <CJC type> = MAN
<A>, <B>, <C> is only used for CUSTOM types for calculating a
compensated thermocouple voltage before converting to temperature:
Ucomp [V] = A + B * Utc[V] + C * Utc[V] * Utc[V].
CH_CONF <RTD type>, <wires>, <Custom Serial. No.>,
<R0/RTPW>, <A/ALR>, <B/BLR>, <C/C1LR>, <C2LR>, <C3LR>,
<C4LR>, <C5LR>,<AHR>, <BHR>, <CHR>, <DHR>, <W660>
(@<ch_list>)
Sets the selected channels for measuring RTD