125614 04
2017-08-11
37
8.0 Errors
Warning
During set-up, make sure that there is correlation between
the software set-up, and the actual distribution of the sensors
connected to the multi-scanner. The software cannot detect
misplacements of sensors.
If the power is lost to the entire test set-up or part of it, it is highly
recommended to restart the calibration.
GENERAL ERRORS
Error description
Solution
LED channel 1 does not flash :
•
Check the power supply
•
Switch the multi-scanner off and on
again
•
If the error still occurs, return the
multi-scanner to the manufacturer for
service
Faulty readout :
•
Check that the connected sensors are
correctly connected and that they are
properly polarised
•
Check that the software is set up
correctly
•
Switch the multi-scanner off and on
again
•
If the error still occurs, return the
multi-scanner to the manufacturer for
service
No contact with the multi-
scanner :
•
Check the power supply
•
Check the RS232 connection. Use
only the supplied cable
•
Switch the multi-scanner off and on
again - and set up the software again
•
If the error still occurs, return the
multi-scanner to the manufacturer for
service