GENERAL ERRORS
Continued
Error description
Solution
When using Hyper Terminal or similar
terminal programs, and communications
fails, check that the program has been set
up correctly.
Faulty readout from
thermocouples:
•
Check that the correct type has been
selected during setup
•
Check that the sensors have been
correctly polarised and connected
•
Check that the correct type of cold
junction has been selected
manually/automatically
•
Check that the multi-scanner is not
subjected to direct or indirect heat
sources
ERRORS – SERIES CONNECTED MULTI-SCANNERS
Error description
Solution
LED’s does not flash
simultaneously on all channel
1’s :
•
Check the RS232 connection. Use
only the supplied cables
•
Switch the multi-scanners off and on
again
Faulty readout :
•
Check that the connected sensors are
correctly connected and that they are
properly polarised
•
Check that the software is set up
correctly
•
Check that analog connection
between the multi-scanners has been
properly established
•
Switch the multi-scanners off and on
again, and set up the software again