For Cold Junction compensation, you will also require a stable
temperature source, for example an ice bath, and a conversion table
to provide the sensor’s corresponding
µ
V value for the bath’s
temperature.
Note…
Before calibrating/adjusting the inputs, the ASM-80x must
have been turned on for at least half an hour.
In order to calibrate the inputs the following calibrated signals must be
used:
Nominal values
Accepted range
•
RTD Input:
50
Ω
(20 to 70
Ω
)
176
Ω
(150 to 200
Ω
)
360
Ω
(330 to 390
Ω
)
1760
Ω
(1500 to 2000
Ω
)
3600
Ω
(3300 to 3900
Ω
)
•
mA inputs:
0 mA
(-1 to +1 mA)
10 mA
(9 to 11 mA)
20 mA
(18 to 22 mA)
•
V Inputs:
0 V
(-1 to +1 V)
5 V
(3 to 7 V)
10 V
(9 to 11 V)
•
mV inputs:
0 mV
(-1 to +1 mV)
35 mV
(25 to 45 mV)
70 mV
(63 to 77 mV)
•
Cold Junction compensation: T-type thermocouple and a
reference junction, for example an
ice bath.