110
23-03-2009
105446 06
THERMAL SPECIFICATIONS –
LIQUID BATHS
ATC-250 A/B
Long term drift (1 year) :
±0.05°C / ±0.09°F
Calibration accuracy (test limit)
±0.13°C / ±0.23°F
Temperature coefficient
5ppm/°C (0-40°C) (32-104°F)
Stability
±0.02°C / ±0.04°F
Reference accuracy
±0.10°C / ±0.18°F
Total accuracy (std. cal. with insert)
±0.50°C / ±0.90°F
Heating time incl. insert
28°C / 82°F to 100°C / 212°F :
3 min.
50°C / 122°F to 100°C / 212°F :
2 min.
100°C / 212°F to 250°C / 482°F :
9 min.
Time to stability
15 min.
Cooling time incl. insert
250°C / 482°F to 100°C / 212°F :
27 min.
100°C / 212°F to 50°C / 122°F :
27 min.
100°C / 212°F to 28°C / 82°F :
55 min.
THERMAL SPECIFICATIONS
ATC-320 A/B
Maximum temperature
320°C / 608°F
Minimum temperature
10°C / 50°F @ ambient temperature 0°C / 32°F
33°C / 91°F @ ambient temperature 23°C / 73.4°F
50°C / 122°F@ ambient temperature 40°C /
104°F