96
23-03-2009
105446 06
THERMAL SPECIFICATIONS
ATC-125 A/B
Long term drift (1 year) :
±0.20°C / ±0.36°F
Calibration accuracy (test limit)
±0.13°C / ±0.23°F
Temperature coefficient
0.005 °C/°C (0-20°C and 26-40°C) (32-68°F and
78.8-104°F)
Stability
±0.03°C / ±0.05°F outside ambient ±5°C / ±9°F
±0.04°C / ±0.07°F ambient ±5°C / ±9°F
Reference accuracy
±0.04°C / ±0.07°F
Total accuracy
±0.30°C / ±0.54°F
Heating time incl. insert
-90°C / -130°F to 23°C / 73.4°F :
15 min.
23°C / 73.4°F to 125°C / 257°F:
13 min.
-90°C / -130°F to 125°C / 257°F :
28 min.
Time to stability
10 min.
Cooling time incl. insert
125°C / 257° to 100°C / 212°F:
12 min.
100°C / 212° to 23°C / 73.4°F:
28 min.
23°C / 73.4°F to -45°C / -49°F:
40 min.
-45°C / -49°F to -80°C / -112°F:
35 min.
-80°C / -112°F to -90°C / -130°F :
30 min.
125°C / 257° to -90°C / -130°F :
130 min.
Refrigerants
R704 (Helium) :
8g.
R1270 (Propylene) :
10g.
THERMAL SPECIFICATIONS –
DRY - BLOCK
ATC-140 A/B
Maximum temperature
140°C / 284°F
Minimum temperature
-35°C / -31°F @ ambient temperature 0°C / 32°F
-20°C / -4°F @ ambient temperature 23°C /
73.4°F
-5°C / -23°F @ ambient temperature 40°C / 104°F