Remove the insertion tube from the calibrator using the tool for
insertion tube supplied with the instrument as shown in Fig. 8.
Caution…
•
The insertion tube must always be removed from the
calibrator after use.
The humidity in the air may cause corrosion oxidation
on
the insertion tube inside the instrument. There is a risk
that the insertion tube may get stuck if this is allowed to
happen.
•
The insertion tube must be removed to avoid damage to
the instrument if the calibrator is to be transported long
distances.
ETC-400 R only
When the ETC-400 R is not in use, it is advisable to store the
calibrator in the carrying case.
Caution – Hot surface
•
Over 50°C/122°F
If the calibrator has been heated up to temperatures
above 50°C/122°F, you must wait until the instrument
reaches a temperature below 50°C/122°F before you
switch it off.
•
You must ensure that the instrument has cooled down to
a temperature below 100°C/212°F before placing it in
the carrying case.