44
2019-04-12
123943
08
8.2
Adjusting and calibrating the instrument
You are advised to return the calibrator to AMETEK Denmark A/S or
an accredited laboratory at least once a year for calibration and
adjustment.
Alternatively, you can calibrate/adjust the calibrator yourself. You will
need a reference thermometer and a reference sensor with a
traceable calibration certificate. Please follow the instructions given
below.
Connect the calibrator to an external precision instrument (e.g. a DTI)
as shown in Fig. 9:
Fig. 9
Hold down the button while pressing the on/off power
control switch.
The instrument is now in adjustment/service mode.
ESC
MENU
Sensor
DTI