If the new coefficients deviate by more than 4% from the
standard values, the instrument will display an ERROR 2 in
the display. The calculated coefficients will be ignored:
Press
to repeat the entire calibration/adjustment
procedure.
Press
when Cont. is flashing to end the
calibration/adjustment procedure and enter a new calibration
date (see section 8.2.1).