Page 9
User Guide
LSP Ex System for Zone 22
Linescanner Head
LSP5 series specifications
LSP5FL
LSP50* LSP52*
Operating wavelength
5µm
Temperature Range
150 to 750°C
500 to 1100°C
302 to 1382°F
932 to 2012°F
System Accuracy
(1)
± 2°C / 3.6°F
(5 to 95% of range)
Focus distance
Fixed at 1200mm / 47in
Target width
Target distance
12mm / 0.5in
(static to 95% radiance)
<1200mm / 47in
Target distance
Field of view 100:1 (0.01 radians)
1200 to 3000mm
47 to 118in
Repeatability
<0.5°C / 0.9°F
Temperature Resolution
(2)
(noise)
<15% of range
<1.5°C / 2.7°F
<2.5°C / 4.5°F
>15% of range
<1°C / 1.8°F
<2.5°C / 4.5°F
Scan Angle
(3)
80°
Scan Rate
Factory set
20Hz
50Hz
50Hz
User adjustable
10 to 100Hz
in 10Hz steps
Speed of response
<5µs
(Tau - linear output on rising edge)
Ambient Stability
<1°/10°
<2°/10°
deltaT[indicated] / deltaT[ambient]
*LSP50 and LSP52 as per LSP5FL unless otherwise stated
(1) Includes LSC Processor error correction
(2) No line averaging. 4LA reduces noise by factor 2
(3) Scan angle may be set in the range 40 to 80° by the LSC-B Processor