Page 32
LSP Ex System for Zone 22
Linescanner Head
User Guide
6 Hotspot Settings
Refer to Section 5 for instructions on how to remove the end plate.
Warning
Isolate the supplies in the safe area before removing the scanner end plate.
Do not open the enclosure or separate the electrical connector when
the equipment is energised, or when an explosive atmosphere may be
present.
Observe precautions for handling electrostatic discharge sensitive devices.
Static-sensitive parts are incorporated in the scanner.
1)
Refer to Fig. 11 for location of the links LK3 and LK4.
2)
Configure link
LK4
for
Peak Temperature
when using with an LMG proces-
sor.
3)
Configure link
LK3
for
Internal Emissivity
.
4)
Set the required Emissivity value using SW1 and SW2 (located next to the
Scan speed switch - see Fig. 11).
Emissivity settings
SW1
SW2
Emissivity
0
0
1.00
9
9
0.99
9
8
0.98
Fig. 11 Location of internal links LK3 and LK4