Specifications
33
Section 5: Specifications
5.1 Specifications
Technology:
RF/ Capacitance
Calibration: None
Modes of Operation: High Level Fail Safe
Repeatability:
2mm (0.08 inch) Conductive Liquids
Response Time:
Less than 1 Second
Time Delay:
0 to 60 Seconds Forward & Reverse Acting
Ambient Electronics: -30 to 70°C (-22 to 158°F) ATEX
-30 to 58°C (-22 to 136°F) IECEx
-40 to 70°C (-40 to 158°F) FM/FMc
Storage Temperature: -40 to 85°C (-40 to 185°F)
Indicators:
LEDs: Green=Power, Red= Relay 1, Red= Relay 2
Self-Check:
Continuous AutoVerify and Manual Certify
Power Supply (Universal Supply):
85 to 250 VAC
21 to 100 VDC Auto-Detecting w/o Jumper Changes
DC to 400 Hz
Power Consumption: 2 Watts Maximum
Relay Contacts:
(2) SPDT
Max Contact Load:
5A/30 VDC/ 5A/250 VAC - Environmentally Sealed
Min Contact Load (DC): 100 mA/12 VDC
0 to 200 mA / 12 VDC Optional
Housing (Electronics): Dual Compartment, Powder-Coated
Aluminum with Two Cable Entries
Cable Entry:
M20 x 1.5 ATEX/IECEx, ¾-Inch NPT FM/FMc
Ingress Protection: IP66 NEMA 4X
Summary of Contents for SIL IntelliPoint RF S*R*L Series
Page 4: ...Contents...
Page 6: ...Section 1...
Page 38: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 32 Section 5...
Page 41: ...Specifications 35 Section 6...
Page 43: ...Control Drawings 37 6 1 FM FMc Control Drawings Continued...
Page 44: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 38 6 1 FM FMc Control Drawings Continued...
Page 45: ...Control Drawings 39 6 1 FM FMc Control Drawings Continued...
Page 46: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 40 6 1 FM FMc Control Drawings Continued...
Page 47: ...Control Drawings 41 6 1 FM FMc Control Drawings Continued...
Page 48: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 42 6 1 FM FMc Control Drawings Continued...
Page 56: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 50 6 2 ATEX Control Drawings...
Page 57: ...Control Drawings 51 6 2 ATEX Control Drawings Continued...
Page 66: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 60 6 4 IECEx Control Drawings...
Page 67: ...Control Drawings 61 6 4 IECEx Control Drawings Continued...
Page 75: ...Control Drawings 69 6 5 Mounting and Wiring for Spark Protector Drawings...
Page 77: ...Control Drawings 71 6 5 Mounting and Wiring for Spark Protector Continued...
Page 78: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 72 6 6 Adding a Padded Capacitor...
Page 79: ...Control Drawings 73 6 6 Adding a Padded Capacitor Continued...
Page 80: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 74 6 6 Adding a Padded Capacitor Continued...
Page 81: ...Control Drawings 75 6 7 Dual Seal Assembly for 700 Series Sensing Elements...
Page 82: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 76 6 7 Dual Seal Assembly Continued...
Page 83: ...Control Drawings 77 6 7 Dual Seal Assembly Continued...
Page 84: ...Appendix A...
Page 88: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual A 4 Appendix B...
Page 90: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual B2 EXIDA Report Results...