SXRXLXX1 - SIL IntelliPoint RF Series User's Manual
A-2
440-0018-731
Sht. 1
of 2
APP'D BY
SGA
ISSUE
EDO NO.
APP'D
DATE
3
2-16-102
THP
2-1-16
4
11-17-100
SGA
11-3-17
EMC Installation Supplement
A.
Purpose:
To provide additional information that is required to be in compliance with
the CE mark of conformity and 2014/30/EU Directive.
B.
Definitions:
1.
I/O Sensor/Measurement/Control Port -- Any port which provides
level measurement, control, and/or DC power.
2.
I/O AC Power -- Any port which provides AC main power to the
instrument.
3.
Housing -- Any enclosure where the sensor and transmitter can be
located.
4.
Non-metallic applications -- any application where the sensor is
not surrounded by a metallic surface.
C.
Installation Specifics:
1.
I/O Sensor/Measurement/Control Ports
Wiring must be twisted pair and run in conduit or an equivalent
shielded environment (i.e. shielded braid, cable, etc.).
The shield terminations must be grounded at the source and
destination ports.
Wiring must be run separate from AC main power and/or any
signal exceeding 75 volts DC or 50 volts AC.
2.
I/O AC Power Port
Wiring must be run either in conduit or an equivalent shielded
environment (i.e. shielded braid, cable, etc.).
The shield terminations must be grounded at the source and
destination ports.
CE Installation Supplement
Summary of Contents for SIL IntelliPoint RF S*R*L Series
Page 4: ...Contents...
Page 6: ...Section 1...
Page 38: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 32 Section 5...
Page 41: ...Specifications 35 Section 6...
Page 43: ...Control Drawings 37 6 1 FM FMc Control Drawings Continued...
Page 44: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 38 6 1 FM FMc Control Drawings Continued...
Page 45: ...Control Drawings 39 6 1 FM FMc Control Drawings Continued...
Page 46: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 40 6 1 FM FMc Control Drawings Continued...
Page 47: ...Control Drawings 41 6 1 FM FMc Control Drawings Continued...
Page 48: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 42 6 1 FM FMc Control Drawings Continued...
Page 56: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 50 6 2 ATEX Control Drawings...
Page 57: ...Control Drawings 51 6 2 ATEX Control Drawings Continued...
Page 66: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 60 6 4 IECEx Control Drawings...
Page 67: ...Control Drawings 61 6 4 IECEx Control Drawings Continued...
Page 75: ...Control Drawings 69 6 5 Mounting and Wiring for Spark Protector Drawings...
Page 77: ...Control Drawings 71 6 5 Mounting and Wiring for Spark Protector Continued...
Page 78: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 72 6 6 Adding a Padded Capacitor...
Page 79: ...Control Drawings 73 6 6 Adding a Padded Capacitor Continued...
Page 80: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 74 6 6 Adding a Padded Capacitor Continued...
Page 81: ...Control Drawings 75 6 7 Dual Seal Assembly for 700 Series Sensing Elements...
Page 82: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual 76 6 7 Dual Seal Assembly Continued...
Page 83: ...Control Drawings 77 6 7 Dual Seal Assembly Continued...
Page 84: ...Appendix A...
Page 88: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual A 4 Appendix B...
Page 90: ...SXRXLXX1 SIL IntelliPoint RF Series User s Manual B2 EXIDA Report Results...