Page 12 of 15
503158 ISSUE 3
EDCR21694
3.2.3 Mounting
The LTX has two mounting holes for a M3.5 screw or #6 ANSI/ASME screw. Do not exceed a tightening force of 2.3 Nm (20 in Lbs). A
suitable thread locking agent is advised.
The LTX has three reference surfaces which may be used to align the head for optimal performance. The primary surface A is formed by
the raised mounts on bottom of the head. The second B is the front of the head and the third closest to the emitter.
When mounting the head the goal is to have the Reference Surface B parallel as possible to the plane of the target. Take care when using
mechanical blocks etc to set up the laser as the window may be scratched.
Measurement Set Up