DC Asterion Series
Overview
M330460-01 Rev A
1-5
Maximum load regulation
+/- 0.02% of rated voltage
Temperature Drift
+/- 100 PPM / degree Celsius
Stability
+/- 0.05% of rated voltage
Constant Current Mode
Maximum line regulation
+/- 0.05% of rated current
Maximum load regulation
+/- 0.15% of rated current
Temperature Drift
+/- 100 PPM / degree Celsius
Stability
+/- 0.05% of rated current
Constant Power Mode
Maximum line regulation
+/- 0.1% of rated power
Temperature Drift
+/- 100 PPM / degree Celsius
Stability
+/- 0.05% of rated power
1.2.6
Output transient specifications
Rated
Voltage (V)
Voltage
Rise Time
(ms), Full
load
Voltage
Rise Time
(ms), No
load
Voltage
Fall Time
(ms), Full
load
Voltage
Fall Time
(ms), No
load
Transient
response
(ms)
Voltage
Ripple/
Noise
RMS, mV
9
Voltage
Ripple/
Noise PK-
PK, mV
10
40 V
60 ms
20 ms
50 ms
800 ms
1 ms
12 mV
75 mV
60 V
60 ms
20 ms
50 ms
900 ms
1 ms
12 mV
75 mV
4)
maximum time, from 0-100% of programming change from zero to rated output voltage with rated resistive load.
Current rise time is same as the voltage rise time
5)
maximum time, from 0-100% of programming change from zero to rated output voltage with No load.
6)
maximum time, from 100%-0 of programming change from rated output voltage to zero with rated resistive load.
Current fall time is same as the voltage fall time
7)
maximum time, from 100%-0 of programming change from rated output voltage to zero with No load
8)
Typical time to recover within 0.75% of rated output voltage for load step change of 90% of rated output current
9)
RMS ripple/noise, over 20 Hz to 300 kHz bandwidth, is measured directly across the output terminals with the
supply operating at full load and nominal AC input line voltage.
10)
PK-PK ripple/noise, over 20 Hz to 20 MHz bandwidth with the supply operating at full load and nominal AC line
voltage.
1.2.6.1
R
EMOTE
S
ENSE
Connection Voltage accuracy/regulation specifications apply at the point where the remote sense leads are
connected.
Line Drop
5%, maximum of rated output voltage per line for models, for 40V to 60V
a larger amount of line drop is allowed, but output voltage regulation specifications no longer apply.
Line Drop
Effect on
Output
Rated output voltage applies at the rear panel output terminals, and line drop voltage subtracts from
the voltage available at the load terminals
CAUTION!
Due to Line Drop Compensation, if Remote Sense is disconnected from unit while the output
is enabled, output voltage will rise a maximum of 10% of
model’s maximum rated voltage,
before faulting
.
Summary of Contents for Sorensen Asterion AST Series
Page 1: ...M330460 01 Rev A www powerandtest com DC Asterion 1U Operation Manual ...
Page 2: ......
Page 26: ......
Page 53: ...DC Asterion Series Overview M330460 01 Rev A 2 27 This page intentionally left blank ...
Page 54: ......
Page 109: ...DC Asterion Series Overview M330460 01 Rev A 3 55 This page intentionally left blank ...
Page 110: ......
Page 112: ......
Page 117: ...DC Asterion Series Index M330460 01 Rev A 3 ...
Page 118: ...Maintenance DC Asterion Series 4 M330460 01 Rev A This page intentionally left blank ...