DC Asterion Series
Overview
M330460-01 Rev A
3-21
3.1.7.3
U
SER
V/I
L
IMITS
The
User V/I Limits
menu allows to set the soft-limits for output voltage and current
to which the unit could be programmed using the front panel or remote digital interface;
default is full scale, refer to Figure 3-51.
Figure 3-51. User V/I Limits Screen
The user limit prevents the supply from being inadvertently programmed above the
user set limit, thus providing a method for protecting the load against damaging
voltages, currents, and/or power levels.
The User V/I Limits menu has the following
fields:
Entry
Description
Voltage
Sets the upper user limit on the programmed output voltage.
Current
Sets the upper user limit on the programmed output current.
Power
Sets the upper user limit on the programmed output power.
3.1.7.4
O
UTPUT
S
ENSE
The
Output Sense
menu allows user to set the output voltage sense of the unit to
either Internal or External, refer to Figure 3-52. When External is selected as output
sense, voltage sense cables must be connected to rear panel of power the supply at
RVS (Remote Voltage Sense) connector.
Figure 3-52. Output Sense Setting Screen
Summary of Contents for Sorensen Asterion AST Series
Page 1: ...M330460 01 Rev A www powerandtest com DC Asterion 1U Operation Manual ...
Page 2: ......
Page 26: ......
Page 53: ...DC Asterion Series Overview M330460 01 Rev A 2 27 This page intentionally left blank ...
Page 54: ......
Page 109: ...DC Asterion Series Overview M330460 01 Rev A 3 55 This page intentionally left blank ...
Page 110: ......
Page 112: ......
Page 117: ...DC Asterion Series Index M330460 01 Rev A 3 ...
Page 118: ...Maintenance DC Asterion Series 4 M330460 01 Rev A This page intentionally left blank ...