Calibration and
Troubleshooting
M370430-01 Rev F
7-9
7
Over Voltage Protection Calibration
Important:
The Voltage Calibration must be done before performing this
procedure. If this is not done the OVP calibration will be inaccurate.
To calibrate the over voltage protection:
1.
Turn on the power supply.
2.
Select the VAP mode to be Off.
3.
Set the voltage set point to be 90% of the model voltage and
the current to be 10% of the model current.
4.
Turn on the main power output.
5.
Execute the OVP calibration SCPI command:
:CAL:PROT
6.
The power supply will enter into self-calibration mode. The
output voltage display blinks OVP CAL. After several
seconds, the calibration will complete and the display will
return to normal operation with output off.
Calibration of the power supply is complete.
Important:
The output (main) voltage must be precalibrated.
The SCPI command (s) for these instructions are:
[:]CALibration[:VOLTage]:PROTection
Summary of Contents for Sorensen XG 100-15
Page 2: ......
Page 93: ...Local Operation M370430 01 Rev F 3 47 3 Figure 3 5 Series Operation ...
Page 99: ...Local Operation M370430 01 Rev F 3 53 3 ...
Page 100: ...Local Operation 3 54 M370430 01 Rev F ...
Page 220: ...Communications Options 6 14 M370430 01 Rev F ...
Page 244: ...Communications Options 6 38 M370430 01 Rev F ...
Page 252: ...Communications Options 6 46 M370430 01 Rev F ...
Page 292: ......
Page 329: ...M370430 01 Rev F A 37 A APPENDIX A SCPI Command Reference ...
Page 330: ......
Page 368: ...APPENDIX D Rack Mount Options and Installation Instructions D 12 M370430 01 Rev F ...
Page 369: ...APPENDIX D Rack Mount Options and Installation Instructions M370430 01 Rev F D 13 ...
Page 370: ......
Page 395: ...Rack Mount Options and Installation Instructions M370430 01 Rev F E 25 ...
Page 396: ......
Page 399: ......