Locking and Unlocking the Front Panel
M370185-01
5-59
5
Setting Dwell Time
The dwell time is the amount of time that is delayed between each
command during the execution of an Auto Sequence program. The dwell
time can be from 0 to 180 seconds and can be changed during the program
execution. The dwell time has a minimum step size of 1 second.
Command:
[:]PROGram[<channel>][:STEP]:DWELl {?|<dwell time> }
Where:
<dwell time > is an integer value from 0 to 180.
For Example:
:PROG:DWEL 12 Results in 12 seconds of wait time after each command
is executed.
Storing an Auto Sequence Program:
Loading a program into memory is done by using the
[:]PROGram[<channel>][:RECord]:STARt
command. After
the start command has been issued, the XTR will record all subsequent
commands into the auto sequence program buffer. The commands will be
recorded into the buffer until the
:PROGram[<channel>][:RECord]:STOP
command is issued.
Commands:
[:]PROGram[<channel>][:RECord]:STARt
[:]PROGram[<channel>][:RECord]:STOP
Important:
If the start recording command is executed then any previously
stored Auto Sequence program is deleted.
For example:
:PROG:STAR
:VOLT 12
:OUTP ON
:PROG:DWEL 60
:OUTP OFF
:OUTP ON
:OUTP OFF
Summary of Contents for Soresen XTR 100-8.5
Page 2: ......
Page 16: ...xiv ...
Page 28: ...1 8 ...
Page 50: ...Local Operation 3 8 M370185 01 Figure 3 3 Front Panel Menu System ...
Page 93: ...Using Multiple Power Supplies M370185 01 3 51 3 ...
Page 94: ...Local Operation 3 52 M370185 01 ...
Page 128: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370185 01 ...
Page 198: ...Remote Operation 5 70 M370185 01 ...
Page 261: ...SCPI Command Tree M370185 01 A 35 A ...
Page 262: ...A 36 ...
Page 279: ...Index M370185 01 IX 3 described 3 7 3 36 number available 3 36 stored values 3 36 ...
Page 280: ...IX 4 ...