Locking and Unlocking the Front Panel
M370185-01
5-61
5
Another benefit to the readback command is it allows the auto sequence
program to be captured and stored on the client side for reloading when
the program is to persist beyond power cycles.
Command:
[:]PROGram[<channel>]:READback?
For Example:
:PROG:START
*CLS
:VOLT 5.4
:CURR 0.25
*IDN?
:PROG:STOP
:PROG:READ?
Response:
*CLS
:VOLT 5.4
:CURR 0.25
*IDN?
:PROG:STOP
Reloading an Auto Sequence Program:
The following procedure will demonstrate how to store a command
beyond a power cycle of the unit.
Important:
This functionality is only available through USB, RS-232 and
RS-485. It is not possible to reload an auto sequence program over the GPIB and
ENET cards.
To reload Auto Sequence program from file:
1. Set the data rate of the device to be 1200 bps.
2. Close the connection to the hyperterminal program and configure the
connection data rate to be 1200 bps.
3. Execute the program recording command:
:PROG:START
Summary of Contents for Soresen XTR 100-8.5
Page 2: ......
Page 16: ...xiv ...
Page 28: ...1 8 ...
Page 50: ...Local Operation 3 8 M370185 01 Figure 3 3 Front Panel Menu System ...
Page 93: ...Using Multiple Power Supplies M370185 01 3 51 3 ...
Page 94: ...Local Operation 3 52 M370185 01 ...
Page 128: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370185 01 ...
Page 198: ...Remote Operation 5 70 M370185 01 ...
Page 261: ...SCPI Command Tree M370185 01 A 35 A ...
Page 262: ...A 36 ...
Page 279: ...Index M370185 01 IX 3 described 3 7 3 36 number available 3 36 stored values 3 36 ...
Page 280: ...IX 4 ...