C-4
M370185-01
Speci
fications
Remote Operation
Table C-2
Remote Operation
Programming Mode
APG
ISOL
Digital
Voltage and Current Output Voltage Programming
0–100%. Voltage control range is 0.0 to 2.0-10.0V in 0.1V increments.
Votage and Current Output Resistive Programming
0-100%. Resistive control range is 0.0 to 2.0 - 10.0K in 0.1K increments.
Output Voltage and Current Monitor
0-100%. Output voltage monitor range is 0.0 to 2.0 - 10.0V in 0.1V increments.
Voltage Programming Accuracy
1
± 0.5% of rated output voltage, max. (0 to 4.0V / 4K Ohm range)
± 0.5% of rated output voltage, typical in other ranges
±0.1% of rated output voltage
Current Programming Accuracy
1
± 1.0% of rated output current, max. (0 to 4.0V / 4K Ohm range)
± 1.0% of rated output current, typical in other ranges
±0.2% of rated output current
Voltage Readback Accuracy
± 1% of rated output voltage
±0.1% of rated output voltage
Current Readback Accuracy
± 1% of rated output current
±0.2% of rated output current
Isolation (Prog and Readback Lines)
With respect to chassis potential:
500 V
With respect to:
chassis potential: 500 V.
negative or positive main output: 1500 V.
Voltage and Current Programming Resolution
0.012% of full scale
Voltage and Current Readback Resolution
Parallel Operation
Up to 4 units in master/slave mode
Series Operation
Up to 2 units (with external
diodes)
Constant Voltage (CV)/Constant Current (CC) Indicator
CV: TTL High (4–5 V)
CC: TTL Low (0–0.6 V)
Shutdown Control
2
Logic low 0.0 - 1.4V
Logic high 2.0 - 15V
Dry contact compatible
OUTPUT ENABLE Auxiliary Control
TTL level or dry contact compatible
Power Supply Status Signal
TTL high: OK (4–5 V)
TTL low: fail (0–0.6 V)
Interlock Enable/Disable
Dry contact. Open/Short: On or Off,
programmable
Summary of Contents for Soresen XTR 100-8.5
Page 2: ......
Page 16: ...xiv ...
Page 28: ...1 8 ...
Page 50: ...Local Operation 3 8 M370185 01 Figure 3 3 Front Panel Menu System ...
Page 93: ...Using Multiple Power Supplies M370185 01 3 51 3 ...
Page 94: ...Local Operation 3 52 M370185 01 ...
Page 128: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370185 01 ...
Page 198: ...Remote Operation 5 70 M370185 01 ...
Page 261: ...SCPI Command Tree M370185 01 A 35 A ...
Page 262: ...A 36 ...
Page 279: ...Index M370185 01 IX 3 described 3 7 3 36 number available 3 36 stored values 3 36 ...
Page 280: ...IX 4 ...