Annunciator
Module
AC Supply
115 to
220 VAC
Power
Supply
L - Live
N - Neutral
G - Ground
G
N
L
Ground
Toggle S
wit
ch
#22 Gauge twisted
pair shielded cable
6"-8
"
[0.1 m - 0.2 m]
supplied
Sensor
Module
DISS
OXY
GEN
Compact
Module
Junction Box
(by others)
Black
Red
Ribbon
Cable
SENSOR
+ -
REMOTE
+ -
COM NC NO
HIGH
ALARM
COM NC NO
LOW
ALARM
Appendix D
Wiring Diagram – Remote Alarm Valve Combo Unit
Hex Plug
NOTE:
1. Do not ground the shield drain wire at sensor or inside the alarm panel back box
2. For multiple sensors, a multi-conductor #22 gauge stranded, shielded and twisted pair
cable ONLY must be used
CAUTION:
To protect from static electricity, ensure to discharge body static before installing the Medical
Gas Alarm and sensors
6" - 8" [0.1m - 0.2 m] #22
gauge stranded, shielded and
twisted pair cable supplied
Marrette the sensor cable in a junction box
(supplied by others) to the installation cable
(supplied by others)
ON
/OFF
#22 gauge stranded, shielded twisted
pair cable only must be used, up to
a distance of 2500 ft [762 m]. In the
presence of any electrical, magnetic
radio, wireless or other interference,
cable installation MUST be placed in a
metallic conduit
Black
Red
SENSOR
+ -
REMOTE
+ -
COM NC NO
HIGH
ALARM
COM NC NO
LOW
ALARM
Black
Red
SENSOR
+ -
REMOTE
+ -
COM NC NO
HIGH
ALARM
COM NC NO
LOW
ALARM
Black
Red
SENSOR
+ -
REMOTE
+ -
COM NC NO
HIGH
ALARM
COM NC NO
LOW
ALARM
AC Supply
115 to 220 VAC
Power
Supply
L - Live
N - Neutral
G - Ground
Toggle Switch
ON/OFF
Ground
G
N
L
6"-8” [0.1 m - 0.2 m] #22 gauge
stranded, shielded twisted pair
cable supplied
Sensor
Module
Sensor
Module
Sensor
Module
Annunciator
Module
www.amico.com