10
Amico Corporation
Compact Display Module - To Change High and Low
Pressure Set Points
To change high and low pressure set points, a dip-switch is located on the back of the display module which is used to
identify the gas of the display module. The dip-switch contains ten switch settings.
PRESSURE ONLY
Factory Default:
• High = 60 Psi, Low = 40 Psi
• Repeat time = 30 min.
HIGH PRESSURE/NITROGEN
Factory Default:
• High = 195 Psi, Low = 140 Psi
• Repeat time = 30 min.
During programming, the “Trend Bar” will Flash!
1. Set switch #6, #7 and #8 to the ON position.
2. The LED will display (HI-), followed by the current set point. Indicating the system is ready to accept a
new High set point. Adjust set point, using the “UP” and “DOWN” push buttons, to the desired value.
3. Set switch #7 to the OFF position.
4. The LED will display (LO-), followed by the current set point. Indicating the system is ready to accept a
new Low set point. Adjust set point, using the “UP” and “DOWN” push buttons, to the desired value.
5. Set switch #8 to the OFF position.
6. The LED will display (I-I-), followed by the current set point. Indicating the system is ready to accept a new
Repeat time set point. Adjust set point using the “UP” and “DOWN” push buttons, to the desired value. [(Display
dd = Disabled) Range from 1 to 60 Minutes]
7. Set switch #6 to the OFF position.
When you have completed step #7, the display module will automatically go into a “RESET” mode. This will store the data
that you had entered.
Compact Alarm Board (v 2.02)
“UP” and “DOWN”
Push Button
Dip-Switch