26
Amico Corporation
WIRING DIAGRAM:
LCD DISPLAY BOARD - REMOTE SENSOR
Note:
For multiple sensors, a
multi-conductor twisted
pair cable can be used.
OXYGEN
# 22 Gauge twisted
pair shielded cable
6”-8” [0.1m-0.2m] supplied
Junction Box
(by others)
Sensor
Module
DISS
Oxygen
Pipeline
Remote location alarm
B4
B3
B2
B1
SE
TUP
CHANGE UP
CHANGE DOWN
SELEC
T/MUTE
AC Supply
115 to 220 V
AC
DC P
ower Cable
:
Connect to
Annunciator Module
G
N
L
L - Live
N - Neutral
G - Ground
Fuse (1 AMPw)
Ground
Toggle S
wit
ch
NOTE:
For multiple sensors, a multi-
conductor #22 gauge stranded,
shielded and twisted pair cable
ONLY must be used
Appendix E
Sensor
Module
6" - 8" [0.1m - 0.2 m]
#22 gauge stranded,
shielded twisted pair
cable supplied
Marrette the sensor
cable in a junction box
(supplied by others) to
the installation cable
(supplied by others)
#22 gauge stranded, shielded twisted pair
cable ONLY must be used, up to a distance
of 1000 ft [304.8 m]. In the presence of any
electrical, magnetic radio, wireless or other
interference, the installation cable MUST
be placed in a metallic conduit
Oxygen
Pipeline
DISS
Red
Black
Orange (15 VDC)
Blue (5 VDC)
Black (Ground)