1.

Remove 3 screws and take the cover off.

2.

Connect the Master Controller input wires 

(Black= Line/White=Neutral) to the line 

voltage using silicone wire nuts.

NOTE:

  Do not turn the power on during the 

wiring work.

3.

Connect the ground wire from the main line 

power to the ground wire in the fixture.

Summary of Contents for LightMaster Multifunction Switch

Page 1: ...Operating and Maintenance Instructions LightMaster Multifunction Switch w w w a m i c o c o m...

Page 2: ...ety Instructions 5 Tool and Material Required 6 Installing the LightMaster Multifunction Switch 7 9 Mounting Drawings 10 11 Wiring Drawing 12 13 Disinfecting Cleaning and Sterilization 14 Technical Da...

Page 3: ...se of the Light Master Multifunction Switch This installation manual gives you instructions on how to install the Light Master Multifunction Switch Please read the following instructions very carefull...

Page 4: ...ith them can lead to injury damage to the luminaire or the equipment In combination with the following signal words the warning symbols means Trademarks All trade names mentioned in these instructions...

Page 5: ...as a low voltage device to ensure the highest levels of safety are met User Profiles No modifications can be carried out by the end user Doing so will void warranty ETL approval and may cause severe r...

Page 6: ...Wire Nuts 4 Single gang electrical box for wall installation Box provided with headwall NOTE Before proceeding with the installation make sure to remove the power to the electrical power supply Light...

Page 7: ...s and take the cover off 2 Connect the Master Controller input wires Black Line White Neutral to the line voltage using silicone wire nuts NOTE Do not turn the power on during the wiring work 3 Connec...

Page 8: ...4 Connect low voltage output wires from Master Controller to the keypad using silicone wire nuts Each keypad will connect to one Red one Blue wire provided and labeled in light fixture Wiring should b...

Page 9: ...ttom terminal block in the keypad in order to operate two functions through the pillow speaker Optional NOTE The nurse call should be connected to only one keypad in the room 8 Once powered the keypad...

Page 10: ...mico Lighting Solutions Mounting Drawings 6 screws can be used to mount the Light Master controller and keypad LightMaster Controller Security Power Unit 6 63 3 82 1 97 0 31 0 31 6 24 3 43 0 17 0 17 1...

Page 11: ...www amico com 11 Mounting Drawings 6 screws can be used to mount the Light Master controller and keypad Key Pad 0 17 0 28 3 28 4 50 2 74...

Page 12: ...12 Amico Lighting Solutions Wiring Drawings NOTE Standard Wiring Diagram is shown Contact factory for custom requirement LightMaster System...

Page 13: ...www amico com 13 NOTE Standard Wiring Diagram is shown Contact factory for custom requirement Wiring Drawings LightMaster System with Secondary Power Unit...

Page 14: ...ation NOTICE Only keypad can be disinfected with the following diluted products Lysoformin Dismozon Hexaquart plus Sagrotan fast disinfecting cleaner NOTICE The local government regulations for disinf...

Page 15: ...olts 0 975 amps Resistive 1 0 amp Ballast at 277 Volts Low Voltage Circuit 5VDC 200 mA Operating Temperature 32 176 F I 0 80 C Master Control Dimensions L 8 X W2 6 X H 1 25 Secondary Control Dimension...

Page 16: ...y part on site or at the factory which is proven to be defective at Amico Lights Corporation s cost Furthermore Amico Lights Corporation will warrant its material to be free from defect for an additio...

Page 17: ...Notes...

Page 18: ...18 Amico Lighting Solutions Notes...

Page 19: ...www amico com 19 Notes...

Page 20: ...R LMS 11 15 19 Amico Lighting Solutions 55 East Wilmot Street Richmond Hill Ontario L4B 1A3 Canada Toll Free Phone 1 877 462 6426 Toll Free Fax 1 866 440 4986 Tel 905 764 0800 Fax 905 763 8587 Email i...

Reviews: