33XR-A Users
Supplement
9/06 1
Change #1
On page
9
, under
General Specification
, change
Operating environment
:
From: 0
º
C to 50
º
C at <70% R.H.
To: 0
º
C to 50
º
C at <70% R.H. for all functions
except 10A ranges
10A ranges: 0
º
C to 40
º
C at <70% R.H.