5
Note:
• DO NOT use solvents to clean the plastic lens.
Case Cleaning: Use soap and water on a damp
sponge or soft cloth.
• DO NOT submerge the unit in water.
TROUBLE SHOOTING
Code
Problem
Action
--- (on display) Target
temperature is
over or under
range
Select target
within
specifications
Battery icon
appears
Possible low
battery
Check and/or
replace battery
Blank display
Possible dead
battery
Check and/or
replace battery
Laser doesn’t
work
Low or dead
battery
Replace battery
CE CERTIFICATION
�
This instrument conforms to the following
standards:
EN 61326-1
-
Electromagnetic Emissions and Susceptibility
EN 61010-1
- General Safety
EN 60825-1
- Laser Safety
Between approximately 250Mhz and 800 Mhz
at 3V/m, the instrument may not meet its stated
accuracy.