27
DC CURRENT (4 to 20mA)
Ranges
0 to 100 %
Accuracy
±0.5 % + 5 dgt
Resolution
0.01 %
Burden voltage
1 mV/mA
Input protection
0.5A/1000V fast blow ceramic fuse
6.3×32mm on μA/mA input
RESISTANCE
Ranges
1000
Ω
, 10 k
Ω
, 100 k
Ω
, 1000 k
Ω
, 10 M
Ω
,
40 M
Ω
(auto/manual ranging)
Resolution
100 m
Ω
Accuracy
±(0.5 % rdg + 8 dgts)
on 1000
Ω
to 1000 k
Ω
ranges
±(1.0 % rdg +10 dgts) on 10 M
Ω
range
±(2.0 % rdg +10 dgts) on 40 M
Ω
range
Open circuit
volts
-0.45 V dc typical
Overload
protection
1000 V dc or 750 V ac rms
CAPACITANCE
Ranges
40 nF, 400 nF, 4 μF, 40 μF, 400 uF
(3999 counts) (auto/manual ranging)
Resolution
0.01 nF
Accuracy
±(3.0 % rdg +10 dgts)
on 40 nF, 400 uF range
±(3.0 % rdg +5 dgts)
on 400 nF to 40 uF range
Test voltange
< 1 V