6) Specifications
General Specifications
Display :
3-3/4 digits 4000 counts LCD display
Polarity :
Automatic
Update Rate :
3 per second nominal
Operating Temperature :
0
°
C to 40
°
C
Relative Humidity :
Maximum relative humidity 80% for temperature up to 31
°
C
decreasing linearly to 50% relative humidity at 40
°
C
Altitude:
Operating below 2000m
Pollution degree:
2
Storage Temperature:
-20
O
C to 60
O
C, < 80% R.H. (with battery removed)
Temperature Coefficient:
nominal 0.15 x (specified accuracy)/
O
C @(0
O
C -18
O
C or
28
O
C -40
O
C), or otherwise specified
Sensing:
Average sensing for AM-100. True RMS for AM-105 TRMS & AM-110 TRMS
Safety:
The meter (all versions) is protected, against the users, by double insulation
per CSA C22.2 No. 1010-1-92, EN61010-1(1995) and IEC61010-1(1995) to CAT II
1000V & CAT III 600V.
Protection-reinforced CE Terminals (to COM) ratings:
V : Category II 1000 Volts AC & DC, and Category III 600 Volts AC & DC.
A / mA
µ
A : Category III 500 Volts AC and 300 Volts DC.
Overload Protections:
Protection-reinforced CE
:
µ
A & mA :
0.63A/500V, IR200kA, F Fuse;
A:
12.5A/500V, IR20kA, F Fuse;
V:
1050Vrms, 1450Vpeak;
Ω
, & Others:
600VDC/VAC rms
Transient protection:
6.5kV (1.2/50
µ
s surge)
Power Supply:
1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2
Power Consumption:
3.2 mA typical
Low Battery
: Below approx. 2.4V
E.M.C.:
Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and
EN61000-4-3(1996)
In an RF field of 3V/m:
Capacitance function is not specified
AC 4.000V range:
Total Accuracy = Specified Ac 700 digits
AC 400.0
µ
A range:
Total Accuracy = Specified Ac 300 digits
Other function ranges:
Total Accuracy = Specified Ac 40 digits
Performance above 3V/m is not specified