19
Precise Low Current DC
(AMP-220 and AMP-320 only)
Range
Accuracy
1) 2) 3)
60.00 A
± (2.0 % + 5 LSD)
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) Specified with DC-zero mode applied to offset the non-zero residual
readings, if any
3) Add 10 LSD to the specified accuracy @ < 4 A
Precise Low Current DC+AC
(AMP-220 and AMP-320 only)
Range
Accuracy
1) 2) 3)
60.00 A
± (2.0 % + 5 LSD)
Frequency:
DC, 50 Hz to 60 Hz
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) Specified with DC-zero mode applied to offset the non-zero residual
readings, if any
3) Add 10 LSD to the specified accuracy @ < 4 A
AC Current
Range
Accuracy
1) 2) 3)
60.00 A
4) 5)
, 600.0 A
± (1.8 % + 5 LSD) @ 50 Hz to 100 Hz
60.00 A
4) 5)
, 600.0 A
± (2.0 % + 5 LSD) @ 100 Hz to 400 Hz
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) AMP-220 and AMP-320: maximum crest factor < 2:1 at full scale & < 4:1
at half scale
3) AMP-210 and AMP-310: specified accuracy is for measurements made at
the jaw center. When the conductor is not positioned at the jaw center,
add 2% to specified accuracy for position errors
4) AMP-220 and AMP-320: add 10 LSD to the specified accuracy @ < 9 A
5) AMP-210 and AMP-310: add 10 LSD to specified accuracy @ < 6 A, and not
specified @ < 0.2 A
DC Current
(AMP-220 and AMP-320 only)
Range
Accuracy
1) 2) 3)
60.00 A
3)
, 600.0 A
± (2.0 % + 5 LSD)
1) Induced error from adjacent current-carrying conductor: < 0.01 A/A
2) Specified with DC-zero mode applied to offset the non-zero residual
readings, if any
3) Add 10 LSD to the specified accuracy @ < 9 A