3
Specifications
General Specifications
Display: 3-1/2 dgt LCD, 1999 counts with function and unit
indicators.
Zero Adjustment: automatic
Low Battery Indication:
�
when battery voltage drops below
operating level.
Polarity indication: automatic, “–” indicated; “+” implied
Display Update Rate, Digits: 2.5/sec, nom.
Operating Temp.: 0 to 50°C, <80% Relative Humidity
Storage Temp: -20°C to 60°C, <70% RH, battery removed
Power Supply: four 1.5V AAA size UM-4 R03 batteries
Battery Life (typical): 200 hrs zinc-carbon.
Dimensions (HxWxD): 170 x 44 x 40 mm, (6.7” x 1.73” x 1.57”)
Weight (including batteries): 250 grams (9 oz)
Housing material: ABS
Input connector: accepts standard miniature thermocouple
connectors (flat blades, spaced 7.9mm, center to center)
Accessories: one type K thermocouple bead probe, soft carrying
case, operator’s manual
Specifications of supplied probe:
1.2m (4 foot) type K thermocouple bead probe (teflon tape
insulated).
Maximum insulation temperature 260°C (500°F). Probe
accuracy ±2.2°C or ±0.75% of reading (whichever is greater)
from 0°C to 800°C.
Agency Approvals:
�
EMC
This product complies with requirements of the European
Community Directive: 89/336/EEC (Electromagnetic
Compatibility).
However, electrical noise or intense electromagnetic fields in
the vicinity of the equipment may disturb the measurement
circuit. Measuring instruments will also respond to unwanted
signals that may be present within the measurement circuit.
Users should exercise care and take appropriate precautions
to avoid misleading results when making measurements in
the presence of electronic interference.
Summary of Contents for TM45A
Page 1: ......
Page 2: ......
Page 3: ...User Manual Bedienungshandbuch Manual de uso Mode d emploi TM45A Digital Thermometer ...
Page 11: ...TM45A Digital Thermometer German Deutsch ...
Page 12: ......
Page 19: ...1 TM45A Digital Thermometer Spanish Español ...
Page 20: ......
Page 27: ...TM45A Digital Thermometer French Francais ...
Page 28: ......