D-1.
General Operation
A diagram and description of the front panel
controls, connection instructions, and the theory
behind resistance measurement is discussed in
this section. This section of the manual contains
complete operating instructions for the
AMPTEC 620A 4R Igniter Tester.
D-2. Front Panel Features and Operation
Power Switch
When the front panel power switch is placed in
the OFF/CHARGE position, all circuitry is
removed from the Igniter Testers panel mount
banana jacks, and the battery pack is connected
to the charging circuit. The battery charger may
then be connected to the 620A-4R charging
jack, in order to facilitate charging the batteries.
For Safety reasons/purposes, when the power
switch is placed in the “ON” position, the
battery pack circuitry is disconnected from the
charging circuit and the front panel charging
jack.
Range Switches
The AMPTEC 620A 4R input range is selected
by depressing the desired range switch
(protected with the given silicone rubber boot
for water resistance) on the front panel. Just
above each resistance range switch (silicone
rubber boot covered) is a red L.E.D. which
illuminates when any one of the four given
ranges is selected. The pushbutton for the (20
Ohm ) lowest resistance range is just below the
display on the left, (see item 11 of the 620A 4R
Front Panel Diagram.). When a given range is
selected an indicator LED informs the user. Also
note that
a resistance range must be selected
after first powering up
the 620A 4R in order to
place it in an
operational mode
, that is when
first turned on a range must first be selected. If
no range is selected upon first “powering up”
the instrument may randomly set itself to a
range that may not be desired until the operator
selects the desired range.
Calibration Access Screws
Around the perimeter of the 620A 4R front
panel you will note there are 10 screws (phillips
head type) that are used for calibration access.
These screws are meant to remain intact and
should only be removed by authorized
personnel (i.e. Calibration Lab staff).
A 0.1 ohm resistor can be used to check the
620A 4R’s zero offset. The AMPTEC 620A
4R’s zero offset can be adjusted while the test
leads are connected to the 0.1 ohm resistor. The
620A 4R zero offset can be adjusted until the
620A 4R display shows 0.1 ohms. (see the
calibration chapter)
Refer the AMPTEC 620A 4R to qualified
service personnel (i.e Calibration Lab) if you
suspect (i.e large error) the 620A 4R
measurement display readings don’t agree with
a test resistors value. Do not attempt to repair
the 620A 4R unless you are qualified.
See Next Page for a Detailed
620A-4R Front Panel Diagram
CHAPTER D - OPERATION, FUNCTIONAL
SELF- TEST AND USE