D-2.3 Main Power Switch 

To turn the unit on, the unit’s main power switch must be pressed for 2 seconds. This 

momentary hold down routine was designed to prevent accidental power on while the 

product is in its carrying case. When the power switch is first pressed and held, the 

instrument’s LCD display should light up, as well as a small LED on the front panel, 

indicating that the unit is now on. If the switch is pressed again, the unit will turn off. The 

main power switch is covered with a silicon rubber boot to provide added water/dust 

resistance. Under the rubber boot, there is a rubber o-ring for added weather 

resistance.  

 

The 620LK meter is set to automatically turn itself off after 5 minutes of no use in order 

to preserve battery life. Contact AMPTEC if you would like to have this feature 

deactivated. 

 

D-2.4 Display and Backlight 

The 620LK’s display backlight is yellow in color, and will light up when the unit is 

powered on. As the 620LK’s display is LCD (liquid crystal display) and high quality, the 

display itself should have a life lasting many years. Should the display become 

damaged or stop operating properly, it can be replaced at an AMPTEC facility. 

 

The unit’s autolock feature turns the meter’s backlight on and will freeze the reading 

unless the “AUTO MEASURE RESET” button on the front panel is pressed or the meter 

senses probe continuity is still established. Additionally, the display backlight is off while 

the measurement portion of its cycle is active. 

 

D-2.5 Resistance Range 

The 620LK can make measurements from 0.00 to 1999.9 m

Ω

The meter starts in Auto Range, but pressing the “AUTO 

MEASURE RESET” button when the meter’s backlight is not on 

allows the user to cycle through the Auto, 200 

mΩ, and 2 Ω 

ranges. 

 

If a measurement is made that results in the unit being over-

range, the display will indicate this with “>>> OR” and “1 - - - -“. 

 

D-2.6 Low Battery Indicator 

If the unit is running low on battery, the display will indicate this by showing “LO BAT” in 

the top left corner. See section E-9.1 on page 17 for information on how to change the 

instrument’s batteries. 

 

D-2.7 Autolock Feature Information 

The 620LK meter has a built-in autolock feature, in which the unit senses continuity was 

established, and automatically samples readings and locks on the most accurate 

resistance measurement.  

 

Summary of Contents for 620LK

Page 1: ...13231 Rooster Springs Rd Austin TX 78737 T 512 858 4045 E info amptec com amptec com AMPTEC RESEARCH MODEL 620LK OPERATION AND MAINTENANCE MANUAL...

Page 2: ...ck Feature Information p 9 D 3 620LK Padded Carrying Case Photo and Information p 11 D 4 Detailed Use Description p 11 SECTION E MAINTENANCE REPAIR AND CALIBRATION E 1 General Maintenance Repair Calib...

Page 3: ...s UAVs drones missiles etc Continuity determinations on fuses electrical cable connections heavy duty metal to metal connections Bonding measurements for metal to metal connections Featuring an all me...

Page 4: ...replacement see section E 9 1 on page 17 4 Perform a visual test lead integrity check It is recommended that the user perform a basic visual inspection of their test lead set to confirm there is no d...

Page 5: ...play is 20 000 counts 4 digit liquid crystal display with a back light that indicates autolock See section D 2 7 on page 9 for additional information about the autolock feature Low Battery Indication...

Page 6: ...a jack test lead set 630 304G 4 4 Wire banana jack test lead set with ground 630 305 4 Twin single banana jack lead set 630 401 4 Single point handheld Kelvin probes 630 405K 10 heavy duty spiked Kelv...

Page 7: ...d start updating the LCD screen once an electrical connection has been made After approximately 3 seconds the meter will produce a beeping sound and the backlight will turn on to illuminate the LCD sc...

Page 8: ...wires then terminate in a variety of configurations depending upon the accessory The 620LK keyed access connector also makes it virtually impossible during normal operation to incorrectly connect lea...

Page 9: ...quid crystal display and high quality the display itself should have a life lasting many years Should the display become damaged or stop operating properly it can be replaced at an AMPTEC facility The...

Page 10: ...ther to produce the final autolocked measurement which is the reading displayed while the backlight is on To assist the user the meter produces an audible beeping indicator along with the lit backligh...

Page 11: ...tem D AUTO MEASURE RESET BUTTON Press and hold to reset a measurement that has been locked Item E Power On Off Switch with LED Press and hold for 2 seconds to power the unit on The display backlight a...

Page 12: ...s the test probes against a conductor meaning continuity is sensed When ready to make another measurement simply place the probes back on the test items and the meter will automatically make another m...

Page 13: ...duct lines and is capable of fixing nearly any issue your instrument may have AMPTEC maintains one sole facility for both the manufacturing and repair of its products AMPTEC RESEARCH CORP 14101 Hwy 29...

Page 14: ...00 mA DC level of test current On a 1 resistor the 620LK 100mA output should get a 100mV drop 0 10V 0 10A For DVM scaling on the 2 range voltage drop is amplified using a x10 for range scaling hence a...

Page 15: ...2 Check for a reading of 0 000 3 Put the output leads across a 0 100 resistance standard 4 The display should read 100 00 m 10 counts depending on the resistor s accuracy If the above test proves unt...

Page 16: ...ated display of the measurement The meter will continue to sample and hold and then resample as long as it senses any continuity Finally connect the meter via the 630 304 test leads across a precision...

Page 17: ...he main PCB back in and replace the front panel and screws E 9 Battery Compartment Design and Description The metal battery compartment is powder coated to make it chemically resistant to battery leak...

Page 18: ...This certification was gained in 2013 and is maintained with quarterly inspections by an independent party AMPTEC also follows UL 913 s quality standards plant wide including the adoption of a qualit...

Page 19: ...____________________________________________________________ ____________________________________________________________ ____________________________________________________________ _________________...

Reviews: