MS322 General Operation and
Application Procedures
General Operation
1. Connect the MS322's ac power cord to the MS322's ac receptacle and plug
the "Hospital Grade" connector to a properly grounded 120Vac, 60Hz
receptacle.
2. Reduce the Low Volt AC Intensity control to the 0/Reset position enabling the
audible "click".
3. Set the Low Volt AC Output Mode control to Tetanize.
4. Set the Pulse/Surge Rate control at the Slow position.
5. Prepare the pad electrodes and apply them to the patient. They may be held
in place by means of retention straps or weight bags.
6. Turn the Power/Timer knob and set it to the desired treatment time.
Note:
When the treatment time is less than ten (10) minutes, turn the Power/
Timer knob clockwise past the ten (10) minute mark and then set it back to
the desired time. The On indicator light (located to the right of the Power/
Timer) will illuminate.
7. Verify that the Low Volt AC Intensity control is set at the 0/Reset position.
8. Slowly increase the Low Volt AC Intensity control to the desired output level.
Note:
If the Reset Intensity indicator light is illuminated, no output will be
obtained until the Intensity control is returned to the 0/Reset position enabling
the audible "click".
MS322 User's Guide
13