Note 

1.    Don’t input voltage and current signal in the “V

Hz” 

jack when measuring the capacitor. 

2.    Be sure to press “REL” to clear zero before measuring. 

3.    Capacitance range only has auto range working mode. 

4.    Release the capacitors completely to avoid damaging 

the meter 

5.    200uF range input reading steady large than 15 Sec. 

6.  Unit: 1uF=1000nF    1nF=1000pF 

Frequency Measurement 

1.  Insert the black leads to “COM” socket and the red 

one to “V

Hz” socket. 

2.    Set the range switch to “Hz” range. 

3.  The range is auto-range. Connect the leads across to 

the signal under measured, the value will be displayed 
on LCD. 

Note 

1.    There is only auto-range at this range. 

2.    When input is over 10V Ac rms, reading is possible, 

but maybe over-range. 

3.  It’s better to use shield cable when measuring small 

signals in noisy place. 

4.    Be careful when measuring high voltage. 

5.    Do not input voltage over 250V DC/AC peak value. 

hFE Measurement 

1.    Set the function switch to hFE range. 

2.  Define the transistor is NPN or PNP type. Insert 

theemitter, base and collector separately to the correct 
hole, then the approx. value will be displayed on 
LCD. 

Diode and Continuity Test 

1.  Insert the black leads to “COM” socket and the red 

one to “V

Hz” socket (the polarity of the red one is 

“+”). 

2.    Set the FUNCTION switch to “

” position. Press 

“DC/AC” key to select diode measuring.   

3.  Forward measurement: connect the red test leads to the 

“+” polarity of the diode under tested and the black 
one to “-”, the approx. forward voltage of this diode 
will be displayed on LCD. 

4.  Backward measurement: connect the red test lead to 

the “-” of the diode and the black one to “+”, and LCD 
will displayed OL. The completely measurement 
includes forward and backward measurement, if the 

result cannot meet with the above, it means the diode 
is workless. 

5.    Diode test include forward measurement and backward 

measurement, if the result is different with above, the 
diode is bad. 

6.    Press “DC/AC” key to select continuity test. 

7.    Continuity test: the buzzer sounds when the resistance 

under measured is less than (70+10) 

Note 

1.

 

  Do not input voltage at this range. 

2.

 

Turn off the power and release all capacitors when    

testing in line. 

Temperature Measurement 

1.    Set the function switch to “ ” range.

 

2.  Insert the cold point of thermocouple too “Temp” 

socket and the working point to the place wanted to 
take temperature, the value will be displayed on LCD. 

Note 

1.  When the input terminal is in open circuit, it will 

display the “normal temp”. 

2.  Do not change the thermocouple, or the accuracy 

cannot be secured. 

3.      Do not input voltage at this range. 

Data Hold 

Press “HOLD” key once, the current data will be hold on 
LCD; press it again, data is canceled. 

Auto Power Off 

1.  When operating after 15 minutes, the meter will be 

auto power off and into sleep mode. Press “POWER 
AUTO OFF” key twice can turn the power on. 

2.    Press “DC/AC” key before turn on, it can cancel auto 

power off function. 

MAINTENANCE 

Do not try to modify the inner circuit. 

Note 

1.    Do not input a voltage over 1000V DC/AC peak value. 

2.  Do not measure voltage at current range, resistance 

range, diode and buzzer range. 

3.    Do not use the meter if the battery is not replaced well 

or the battery case is not fixed. 

4.  Before replacing battery or fuse, release the test leads 

from the test point and turn power off. 

Summary of Contents for VC97

Page 1: ...nce frequency temperature etc DESCRIPTION VC97 is a high performance high accuracy 3 3 4 digit 32mm digit high by LCD displaying meter The function includes DCV ACV DCA ACA resistance capacitance freq...

Page 2: ...rature 10 50 Relative humidity 80 RH Power 2pcs 1 5V AAA 7 battery Dimension 185 L 93 W 35 H mm Weight Approx 290g including battery Accessories Manual gift box holster TP01 temperature probe test lea...

Page 3: ...0A less than 15sec Overload Protection 0 5A 250V fuse 13A 250V fuse Frequency Response 40 100Hz under 10A range 40 400Hz at other range Table 6 Resistance Range Accuracy Resolution 400 0 8 5 0 1 4k 0...

Page 4: ...5 4 0 001Hz 100Hz 0 01Hz 1000Hz 0 1Hz 10kHz 1Hz 100kHz 10Hz 1MHz 100Hz 30MHz 1kHz Input Sensor 0 7V Overload Protection 250V DC AC peak value Table 9 hFE Range Accuracy Test condition hFE NPN or PNP 0...

Page 5: ...ange key Select auto range or manual range 2 6 DC AC key Select DC AC mode 3 Knobe Change measuring function and range 4 Temperature jack 5 Voltage Current Resistance Frequency GND jack 6 Battery case...

Page 6: ...on switch to A range press DC AC key to select AC working mode connect the test leads across to the circuit under tested the current value and polarity of the point which red lead connect will display...

Page 7: ...and the black one to the approx forward voltage of this diode will be displayed on LCD 4 Backward measurement connect the red test lead to the of the diode and the black one to and LCD will displayed...

Page 8: ...he circuit boards and case and can cause the multimeter to work improperly although the holster can provide enough protection 8 Wipe the multimeter with a damp cloth occasionally to keep it looking ne...

Page 9: ...ers of each device is not necessarily performed except those mandated by government requirements 3 Customers are responsible for their applications using ATI components In order to minimize risks asso...

Reviews: