Installation Procedure for bkSpecto
IG42955-03
16-01724 rev 00
1/16
Installation Procedure for bkSpecto
IMPORTANT:
This procedure must be carried out only by BK Ultrasound service personnel or their authorized
representatives. The procedure covers unpacking and set-up of the system including connecting to the
power outlet and a possible network connection.
All scanners leaving the factory are configured for mains voltage 100-240V 50/60Hz and English
language. The mains voltage does not have to be changed as the power supply accepts 100-240V
50/60Hz.
CONTENTS
PAGE
1
Necessary Tools
-
Small thin slotted screw driver (e.g. 2mm or 3/32 in).
-
Tools for electrical safety test.
-
Torx #10 and #20 for printer set up.
-
Torx #25 for transducer holder frame and additional transducer holders.
-
Checking Procedure available in user guide.