Installation Procedure for bkSpecto
IG42955-03
16-01724 rev 00
3/16
10. Remove the scanner from the pallet.
11. Empty the box from step 6.
12. Unpack the transducer holder frames from the
bags.
Note:
The transducer holder frame package contains: 3 transducer holders, 1 gel holder, 2 transducer holder frame
trays, and four screws.
3
Check the fuses
1. The fuses are placed under the engine next to the power inlet. Check the fuses are correct:
Scanner Overview
Procedure
Press the black latch
(listen for a click sound)
and remove the fuse
holder.
Check the correct
fuses are used:
T4AH250V
Insert the fuses into the holder and insert the
holder into the engine again.
CAUTION:
While inserting the fuse holder a click sound will indicate that it is mounted properly.