18
-
Measure it and remove the transducer
Sample piece
Measured value
Couplant
4-3-4. Two point calibration for Sound Velocities
Prepare a sample piece "or two pieces" whose thickness are
known.
Step
1
. Perform the Probe-Zero Function.
Perform the "
4-2-2. Probe - Zero Function
".
Step 2.
Measure the Thickness of the First Point.
Apply couplant to the first point of the
sample piece. Press the transducer
against it. The thickness value of the
first point is displayed.
Remove the transducer from the
surface after the measurement.
Step 3.
Input the Value of the First Point.
Press
the switch. The mm blinks.
Input the value of the first point using
the
switch and
switch.
Press
the switch to display
of next step.
Step 4.
Measure the Thickness of the Second Point.
Apply couplant to the second point of
the sample piece. Press the transducer
against it. The thickness value of the
second point is displayed.
Remove the transducer from the
surface after the measurement.
or
Corrected value
Proceed next step
Measure it and remove the transducer
Measured value