29
-
6. Specifications
Measurements
Measurement method
.............................
Ultrasonic pulse echo technique
Unit
.........................................................................................................................
mm, m/s
Sampling rate
Single point mode
...........................................................................
0
7
times/sec.
Scan
mode
........................................................................................
16
times/sec.
Measurement range
Pulse
-
echo mode
7
MHz
transducer AD
-
3255
-
02
.........................
0.8
to
200.0
mm
5
MHz
transducer AD
-
3255
-
03
.........................
1.0
to
300.0
mm
5
MHz
transducer AD
-
3255
-
04
.........................
2.0
to
300.0
mm
Echo
-
echo mode
5
MHz
transducer AD
-
3255
-
04
.........................
4.0
to
30.0
mm
Measurement accuracy
Range
0
0.8
to
50.0
mm
......................................
±
0.1
mm
±
0.5
%
50.1
to
300.0
mm
......................................
±
0.1
mm
±
1.5
%
The operation temperature range
...................................................
5
to
40
°C
The range of the sound velocity
...................................
1000
to
9999
m/s
Display range
............................................................................
0.65
to
600.0
mm
Display resolution
..............................................................................
0.1
/
0.01
mm
The accuracy depends upon the measurement conditions
and the surface of the material.