23
9-2. Data format
S T
,
+
0
0
0
0
0
.
0
0
g C
R
L
F
Header
Data
Unit Terminator
There are four types of headers:
ST : Stable weight data (including % data)
QT : Stable counting data
US : Unstable weight data (including count and %)
OL : Out of weighing range (Over)
The data is normally 9 digits including a decimal point and a sign.
There are 11 types of units:
g : Weight data “gram”
P C : Counting data “pcs”
% : Percentage data “%”
o z : Weight data “decimal ounce”
l b : Weight data “decimal pound”
o z t : Weight data “troy ounce”
c t : Weight data “carat”
m o m : Weight data “momme”
d w t : Weight data “penny weight”
G N : Weight data “grain”
t l : Weight data “tael”
The terminator is always C
R
L
F
.
Example of output data:
Weight data “gram”
S T
,
+ 0 0 1 2 3 4
.
5
g
C
R
L
F
Counting data
Q T ,
+ 0 0 0 1 2 3 4 5
P C
C
R
L
F
Percentage data
S T
,
+ 0 0 0 1 2 3
. 4
%
C
R
L
F
Out of range “gram” (+)
O L
,
+ 9 9 9 9 9 9
. 9
g
C
R
L
F
Out of range “pcs” (-)
O L
,
- 9 9 9 9 9 9 9 9
P C
C
R
L
F
9-3. Data output mode
Stream mode
Set the function “
prt 0
”.
The balance outputs the current display data. The data-update rate is approximately
10 times per second. This rate is the same as the display-update.
The balance does not output data while it is in the setting mode.
Separator
(“ ” shows a space.)