25
10. ID NUMBER AND GLP
The ID number is used to identify the balance when Good Laboratory Practice (GLP)
is used. The following GLP data is transmitted to an AD-8121 printer or a computer
using the RS-232C interface.
The result of calibration (“Calibration report”)
The result of calibration test (“Calibration test report”)
The “Start block” and “End block” for GLP data
10-1. Setting the ID number
1. Press and hold the SAMPLE key to display
func
.
2. Press the SAMPLE key several times to display
id
.
3. Press the PRINT key. Enter the ID number using
the following keys.
SAMPLE key To select the digit blinking to
change.
RE-ZERO key To set the character of the digit
selected. See the table below for
the “display character set”.
PRINT key
To store the value and proceed to
the next step.
MODE key
To cancel the value and proceed
to the next step.
4. When the above operation has completed,
func
appears after
end
.
5. Press the MODE key to return to the weighing
mode.
Display character set
0 1 2 3 4 5 6 7 8 9 - A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
0 1 2 3 4 5 6 7 8 9 - a b C d e f g H i j k l m n o p q r s t U v w x y z
“ ” : Space
Press
several times
func
id
000000
end
func
Set using the
relevant keys
Returns to
the weighing mode
Press
several times
Press and hold