14
8.3. Readability
Mass can be displayed on FG-CWP series with the following four types of readability (scale
interval or "division"). The ratio of readability and weighing capacity expressed as display
resolution is shown in the following table.
Function table
Display resolution
re5o 0
(low resolution)
1/3000
re5o 1
(normal resolution)
1/6000 or 1/7500
re5o 2
(high resolution)
1/12000 or 1/15000
re5o 3
(max resolution)
1/30000
Readability is set to normal resolution at time of shipping. It can be changed in the function table
item
re5o
as needed for the usage environment. See "
" for details on how to
configure the settings.
□
Calculation to display the count in counting mode is done at max resolution regardless
of the set display resolution.
9. Counting Function
This function stores the mass (unit weight) of a single item from a known number of sample masses,
and calculates and displays the number of items based on that value and the total mass.
□
The unit mass is stored in non-volatile memory
even when power is turned off.
1. Press the MODE key to display the weighing
unit “pcs”.
(pcs = pieces)
2. Display the unit mass stored by pressing the
SAMPLE key. The figure on the left is the
sample count.
3. Press the PRINT key to change the sample
count. Select from 5, l0, 20, 50 and l00 pcs.
4. When “-“ appears at the right side of the number
of samples, press the ZERO key to zero the
scale. If necessary, place a container on the
weighing pan, and press the TARE key.
Confirm that the right side of the number of
samples shows zero.
5. Place the number of samples shown on the
weighing pan or into the container.
Press
(select)
ZERO
STABLE
ZERO
STABLE
Number of samples
STABLE
Container
Weighing
Pan
ZERO
NET
STABLE
Sample
Press
Press
(select)
Press Confirm
zero
STABLE
Weighing