15
6. Confirm that the stability indicator is lit, and then press
the SAMPLE key to calculate and store the unit
mass. The display changes to the count using the unit
mass stored. Remove the sample.
□
The mass of the number of samples must be at least
the values below, regardless of the sample count.
Weighing capacity
Minimum
mass
FG-30KCMWP
10g
FG-60KCLWP
20g
FG-150KCLWP
50g
If the sample mass is small,
lo ut
is displayed and returns to the display in step 5.
Increase
the number of samples (see step
3
) and try again.
□
If you press the SAMPLE key without reaching the minimum mass in the table above,
you will be returned to weighing mode.
10. Comparator Function
The comparator function compares the weighed value against the set values and displays the results
using LEDs (yellow / green / red).
Comparison is done based on the following table, and the
results are displayed/output.
Judgment
result
Judgment formula
HI
HI limit value < Displayed value (Or when positive value)
OK
LO limit value
≦
Displayed value
≦
HI limit value
LO
Displayed value < LO limit value (Or when negative value)
□
Select the comparison condition in function table item
Cp
. (Refer to
)
Function
table
Content
Cp 0
No comparison (comparator function unavailable).
Cp 1
All weighed values are compared regardless of stability.
Cp 2
All stable weighed values are compared.
Cp 3
Weighed values except those at -4d to +4d are compared regardless of
stability
Cp 4
All stable weighed values except those at -4d to +4d are compared.
Cp 5
Weighed 5d and above are compared regardless of stability.
Cp 6
Stable weighed values at +5d and above are compared.
d= readability (scale interval or "division") (see the “basic manual” included with the product)
Even in the counting function, “d” is the readability in the mass display.
□
The HI/LO values are stored in non-volatile memory even when power is turned off.
□
The HI/LO values are the same in both mass display and count display.
□
The decimal point is ignored for HI/LO values.
When parameter is “0010000” with the FG-30KCMWP (decimal point ignored)
Yellow
Green
Red
LED Display
LO OK HI
STABLE
STABLE
Press