Operator’s
Manual
Rev
E
4/2/12
pg.
17
Periodic
Maintenance
Cleaning
the
Extraction
Vessel
Some compounds extracted during the fat extraction procedure can create a black residue in the bottom of the
Extraction Vessel. The vessel can most easily be cleaned by using a drill with a long shafted stiff wire brush. This
treatment will remove a lot of the black coating that accumulates on the bottom of the vessel. When the vessel
bottom is clean, put some solvent on a cloth or paper towel and wipe down the inside of the vessel in an effort to
remove any oil residue and particles that have been loosened from the vessel bottom.
IMPORTANT
Do not use soap and/or water to clean the vessel.
When cleaning the vessel, do not damage the level sensor tip. The tip may be cleaned
with window cleaner or alcohol.
Cleaning
the
exterior
of
the
instrument
Wipe the outside of the instrument with window cleaner or a damp sponge as necessary.
Do not use Acetone to clean the instrument.
Cleaning
the
solvent
filter
If filter bags are not sealed properly and sample is lost from the bag, it will be recovered in the solvent filter in the
exhaust line.
To clean the solvent filter, follow the procedure detailed below.
1.
Remove the rear cabinet of the instrument.
Rear
View
of
instrument
with
cabinet
back
removed
2.
Unscrew the compression fitting.
3.
Unscrew the filter body.
4.
Unscrew and clean the metal filter screen.
5.
Reassemble the filter and compression
fitting. Make sure the valve body is tight and
the Teflon washer has not come out of the
filter body.
Compression
Fitting
Filter
body
Summary of Contents for XT10
Page 2: ...This page intentionally left blank ...
Page 4: ...This page intentionally left blank ...
Page 25: ...Operator s Manual Rev E 4 2 12 pg 25 Appendix C Electrical Diagram pg 1 of 3 ...
Page 26: ...Operator s Manual pg 26 Rev E 4 2 12 Appendix C Electrical Diagram pg 2 of 3 ...
Page 27: ...Operator s Manual Rev E 4 2 12 pg 27 Appendix C Electrical Diagram pg 3 of 3 ...