Chapter 3 — Performance Verification
3-7 Spurious Signals Tests
MG3702xA MM
PN: 10370-10372 Rev. B
3-7
Harmonics and Harmonically Related Signals Test
Harmonic Test Setup
Connect the equipment shown in
, as follows:
1.
Connect a BNC 6 dB pad to the MG3702xA rear panel 10 MHz REF OUT port.
2.
Connect a BNC cable to the 6 dB pad on the MG3702xA rear panel 10 MHz REF OUT to the
spectrum analyzer's external reference input.
3.
Connect the MG3702xA RF Output to the spectrum analyzer RF input.
4.
Set up the spectrum analyzer as follows:
a.
Press the
PRESET
key.
b.
Press
AUX CTRL
.
c.
Press
Rear Panel
.
d.
Press
10MHz
and set to “EXT”.
Harmonic Test Procedure
1.
Determine the first test frequency to test based on the test record, options and model number.
2.
Set up the MG3702xA as follows:
a.
Preset the instrument by pressing the
Preset
button. The CW menu is displayed.
b.
Press
Edit Frequency
to open the current frequency parameter for editing and set
F0
to the
first appropriate test frequency (based on model and options) in the test record.
c.
Press
Edit Level
to open the current power level parameter for editing.
d.
Set
L0
to +10 dBm or to the maximum specified power level, whichever is less. Refer to
“MG3702xA Technical Data Sheet”
for the maximum specified power levels.
For the electronic version, refer to the MG3702xA Technical Data Sheet, PN: 11410-00429
on the Anritsu Internet site,
http://www.anritsu.com.
3.
Set up the spectrum analyzer as follows:
a.
Press the
AMPLITUDE
key. Press
REF LVL
and enter the current power level setting (L0) of
the MG3702xA.
b.
Press the
FREQUENCY
key and enter the current frequency setting (F0) of the MG3702xA.
c.
Press the
BW
key and press
MAN
. Set to the first appropriate test frequency RBW/VBW
value (based on model and options) shown in the test record.
d.
Press the
SPAN
key and set to the first appropriate test frequency SPAN value (based on
model and options) shown in the test record.
4.
Press the
PEAK SEARCH
key, then select
MARKER DELTA
.
5.
Press the
FREQUENCY
key on the spectrum analyzer and enter next harmonic frequency listed in
the test record.
6.
Enter the reading from the spectrum analyzer in
,
Harmonically Related Signals Test (without Option 15)”
7.
Repeat
through
for each of the harmonic frequency listed in the test record.
Note
Power line and fan rotation spurious emissions are tested as part of the single sideband phase noise
test in
.
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...
Page 183: ......