3-7 Spurious Signals Tests
Chapter 3 — Performance Verification
3-8
PN: 10370-10372 Rev. B
MG3702xA MM
8.
On the MG3702xA, press
Edit Frequency
to open the current frequency parameter for editing and
set F0 to the next appropriate test frequency (based on model and options) shown in the test
record.
9.
Determine if the power level for the MG3702xA and spectrum analyzer changes based on the new
frequency or options and if so, adjust the MG3702xA L0 and spectrum analyzer amplitude as
needed.
10.
Press the
SPAN
key and enter the value listed in the test record for the appropriate test frequency.
11.
Press the
BW
key and enter the value listed in the test record for the appropriate test frequency.
12.
On the MG3702xA, press
Edit Frequency
to open the current frequency parameter for editing and
set F0 to the next appropriate test frequency (based on model and options) shown in the test
record.
13.
Repeat
for each of the test frequencies listed in the test record.
Non-Harmonic Signals Test
For the following test, refer to the test record tables in
“Section 3-7 Non-Harmonic Signals Test
Non-Harmonic Test Setup
Connect the equipment, shown in
1.
Connect a BNC 6 dB pad to the MG3702xA rear panel
10 MHz REF OUT
port.
2.
Connect a BNC cable to the 6 dB pad on the MG3702xA rear panel 10 MHz REF OUT to the
spectrum analyzer's
EXT REF IN
port.
3.
Connect the MG3702xA
RF Output
to the spectrum analyzer’s
RF Input
.
4.
Set up the spectrum analyzer as follows:
a.
Press the
PRESET
key.
b.
Press
AUX CTRL
.
c.
Press
Rear Panel
.
d.
Press
10MHz
and set to “
EXT
”.
Non-Harmonic Test Procedure: Part 1
For the following test, refer to the test record table in
,
“Section 3-7 Non-Harmonic Signals
“Section 3-7 Non-Harmonic Signals Test (Part 1)”
1.
Set up the MG3702xA as follows:
a.
Preset the instrument by pressing the
Preset
button. The CW menu is displayed.
b.
Press
Edit Frequency
to open the current frequency parameter for editing and set
F0
to the
first appropriate test frequency (based on model and options) in the test record.
c.
Press
Edit Level
to open the current power level parameter for editing.
d.
Set
L0
to +10 dBm or to the maximum specified power level, whichever is less. Refer to
“MG3702xA Technical Data Sheet”
for the maximum specified power levels.
For the electronic version, refer to the MG3702xA Technical Data Sheet, PN: 11410-00429
on the Anritsu Internet site,
http://www.anritsu.com.
Note
Power line and fan rotation spurious emissions are tested as part of the single sideband phase noise
test in
Summary of Contents for MG3702xA Series
Page 4: ......
Page 5: ......
Page 8: ...TG 8 PN 10370 10372 Rev B MG3702xA MM ...
Page 24: ...1 12 Test Equipment List Chapter 1 General Information 1 10 PN 10370 10372 Rev B MG3702xA MM ...
Page 42: ...2 5 RF Deck Assemblies Chapter 2 Functional Description 2 18 PN 10370 10372 Rev B MG3702xA MM ...
Page 120: ...5 5 Troubleshooting Tables Chapter 5 Troubleshooting 5 24 PN 10370 10372 Rev B MG3702xA MM ...
Page 182: ...Index 4 PN 10370 10372 Rev B MG3702xA MM ...