103
Appendix B Troubleshooting
Following error message is appeared at the power on.
“Error found in back-up data. All settings and measured data cleared to default.”
If the above error message is appeared after power on again, the internal-memory back up battery life
is ended. Please, call your nearest Anritsu sales office or the representative to replace the battery
with a new one by our service department.
Note: The battery life of this instrument is 7 years after delivery. Early replacement is recommended.
The instrument is loaded with the Microsoft® MS-DOS® Operating System and Microsoft® Win-
dows® Operating System, so you can run applications or make Windows® setup changes the
same way as you run Windows® on a personal computer.
This appendix lists problems that may result from Windows® setup changes and the suggested
remedies.
The touch screen does not work
The touch screen would be disabled if you alter the mouse driver setting with the Windows Setup
program. Set the mouse to TTwin v3.20:AR5000 and Digitouch with the Windows Setup pro-
gram and restart Windows®. The touch panel will become operable after Windows® has been
restarted.
The front-panel LCD display is off
The front-panel LCD would be turned off if you select ChipsCPL from Windows® Control Panel,
start the display Driver Control Panel program, and then select CRT in the Display group box.
If an external display is connected to the instrument, select Both in the Display group box while
observing the external display. The LCD display will be turned on. If an external display is
not connected to the instrument, turn off the power, but not before checking to see that the hard
disk access lamp is off. The LCD display will be restored when the instrument is turned on
subsequently.
The external display is off
The external display would be turned off if you select ChipsCPL from Windows® Control Panel,
start the display Driver Control Panel program, and then select LCD in the Display group box.
Select Both in the Display group box while observing the front-panel LCD. The external display
will be restored.
Application characters are not displayed correctly
Application characters would not be displayed correctly if you delete the font of characters used
by the instrument application by running the Fonts program from Windows® Control Panel.
Reinstall the font deleted by running the Fonts program again.
Summary of Contents for MP1632C
Page 17: ...xvi...
Page 22: ...1 CHAPTER 1 OVERVIEW...
Page 30: ...9 CHAPTER 2 SETUP...
Page 35: ...14 Blank...
Page 36: ...15 CHAPTER 3 PANEL DESCRIPTION AND CONNECTIONS...
Page 45: ...24 Blank...
Page 46: ...25 CHAPTER4 STARTING AND STOPPING THE MP1632C...
Page 54: ...33 z Group box A frame or box that encloses a set of related controls...
Page 55: ...34 Blank...
Page 56: ...35 CHAPTER 5 DISPLAY DESCRIPTION...
Page 75: ...54 Blank...
Page 76: ...55 CHAPTER 6 OPERATING INSTRUCTIONS...
Page 114: ...93 CHAPTER 7 PERFORMANCE TEST...
Page 117: ...96 Blank...
Page 118: ...97 CHAPTER 8 MAINTENANCE SCRAPPING...
Page 121: ...100 Blank...
Page 122: ...101 APPENDIXES...
Page 127: ...106 Blank...