3 Installation
14
C58IB004EN-D
3 Installation
3.1 Installation Requirements
3.1.1 X-Ray Diffractometer Requirements
3.1.2 Mounting Adapter
On certain goniometer supports an adapter is nec-
essary to mount DHS 1100 to the goniometer.
3.1.3 Electrical Requirements CCU 100
• Mains voltage: AC 100 - 240 V
• Mains frequency: 50 or 60 Hz.
• mains connector (1): IEC 60320 C13
• mains fuses: 2 x T 6.3 A H 5 x 20 mm (ceramic
tube)
• Remote control interface: RS 232 C
• RS 232 connector on CCU (2): D-Sub DE-9
(male)
• Cable type: null modem
3.1.4 Compressed-Air Cooling
For operation of DHS 1100 with compressed-air
cooling, the main requirements for the compressed-
air supply are:
• Cooling air pressure: min. 2 bar rel. / max. 4 bar
rel. (rel. = relative to atmospheric pressure).
• Air flow rate at 2 bar: 0.12 m³/min
• The air must be free of oil and dust
3.1.5 Vacuum Equipment Requirements
(Optional)
Vacuum equipment requirements:
• Pumping speed acc. DIN 28400:
4.2 m
3
/h
• Ultimate partial pressure acc. DIN 28400:
1E-4 mbar
• Ultimate total pressure acc. DIN 28400:
2E-3 mbar
• Vacuum connector on sample chamber:
DN16 ISO-KF
3.2 Unpacking the Instrument
TIP:
Anton Paar GmbH gives no warranty for the
dome if it is not handled according to our instruc-
tions.
If the dome is already broken when unpacking DHS
1100, please take pictures that show explicit the
broken dome with the original packing material. Oth-
erwise no warranty can be accepted by Anton Paar.
CAUTION
Radiation
The DHS 1100 represents an open system and
therefore is only allowed to be operated on goni-
ometers equipped with a radiation enclosure
NOTICE
We recommend to use the Air Service Unit by
Anton Paar GmbH (refer to the reference guide
C58IB002EN).
NOTICE
Risk of damage
• Unpack the system with care.
• Do not drop the instrument.
• The graphite dome is a fragile item, handle it
with particular care.
• Always lift the dome by the fixing ring, do not
touch the graphite dome.