7 Troubleshooting
C58IB004EN-D
31
7 Troubleshooting
7.1 Error Messages
* not applicable for DHS 1100 (valid for other non
ambient attachments)
7.2 General Messages on Display
7.3 Error Analysis
For information about the error analysis refer to the
reference manual C58IB002EN.
Alarm
Message
Alarm
Name
Description
E01
Sensor
Break
The electronic circuit
of the temperature
sensor is interrupted.
E02
Loop
Break
The electronic circuit
of the heater is inter-
rupted.
E03
No Water
Flow
There is no or too little
water flow.*
E04
Housing
Overtem-
perature
The protective ther-
moswitch in the sam-
ple stage housing is
activated.
E05
Air Cool-
ing Fail-
ure
Air cooling flow is off
or insufficient.
E06
sample
plate not
mounted
The contact switch
does not detect the
sample plate.*
E07
Chamber
not closed
The contact switch
does not detect the lid
of the instrument.*
E08
Current
cable dis-
connected
The contact switch
does not detect the
current cable.*
E09
CCU 100
Air Cool-
ing Fail
The protective
thermoswitch inside
CCU 100 is activated.
E10
Other
Messages
Internal errors of
CCU 100
contact
Anton Paar GmbH.
E11
Sensor
Spread
Failure
Activated in case of
too big temperature
difference between
the sensors (sample
holder and heater).*
E12
Flow-Con-
trol Exten-
sion
missing
Activated in case of
missing flow control-
ler.*
E13
Power
Stage
Fault
Contact Anton Paar
GmbH
E98
No /
Unknown
Instru-
ment
present
Activated in case that
there is no connec-
tion between chamber
and CCU.
E99
No /
Unknown
module
present
Activated in case that
there is no power
module present.
Message
Description
8888
Normal during booting. If the mes-
sage doesn´t disappear after boot-
ing, the power module doesn´t boot
properly. Check the error code.
9999
This value is shown in case of sen-
sor break. Check the error code.
Alarm
Message
Alarm
Name
Description