4 Installation
B40IB006EN-D
19
4.4.1 Power Supply for the Heater
Power for the heater of the HTK 1200N is supplied
with two high current cables. Plug the cable for the
heater to the heater connector on the rear side of
the CCU 1000 (Fig. 10 (2)). Connect these cables
with the two high current cables at the lid of the
HTK 1200N as described below:
Connecting procedure:
Disconnecting procedure:
4.4.2 Thermocouple of Stationary Sample
Holder and Sample Spinner
For correct installation please follow this instruction:
Stick the plug-in connection (1) of the cable for sam-
ple holder flange into the locknut (2) of the sample
holder at the bottom of the sample holder flange.
Then connect the other end of the cable to the con-
nector "Temperature Sensor" on the connector box
(Fig. 8 (3).
WARNING
Risk of injury
This instrument must be connected to protective
ground.
CAUTION
Risk of damage
Switch off the CCU 1000 and disconnect the
cable for the mains supply, before disconnecting
and connecting the high current cables and the
thermocouple cable.
Connected
Disconnected
1
2