Turn off the suction pump and disconnect suction pump and all ancillary components from the
endoscope.
Dry the surface of the endoscope, the air/water valve(inapplicable to VET-3512), the suction valve and the
biopsy valve with a sterile lint-free cloth.
4.5.8
Drying
Place the endoscope, the air/water valve(inapplicable to VET-3512), the suction valve and the biopsy
valve on the sterile towel; replace sterile towel every 4hours.
Block the suction cylinder, air/water cylinder, instrument channel inlet with the channel plug; block the
air/water supply connector with sealing plug, connect the injection tube to the endoscope connector.
Use a 50ml syringe to respectively inject 75%~95% alcohol into the air/water cylinder (inapplicable to VET-
3512)and suction cylinder through the injection tube at least 15 times (750ml).
Use a 50ml syringe to respectively inject air into the air/water cylinder (inapplicable to VET-3512)and
suction cylinder through the injection tube at least 15 times (750ml).
Dry the outer surface of the endoscope, the air/water valve(inapplicable to VET-3512), the suction valve
and the biopsy valve with a sterile lint-free cloth.
Dry the inner surface of air/water cylinder(inapplicable to VET-3512), suction cylinder and instrument
channel inlet with sterile cotton swab.
Install the water/air valve(inapplicable to VET-3512), suction valve and the biopsy valve on the endoscope.
4.5.9
Other accessories cleaning, disinfection and sterilization
Refer to respective instruction manuals of accessories to conduct cleaning and disinfection procedures.
Ultrasound cleaning is the ideal method to implement the particle-level physical cleaning. If possible,
sterilize these accessories with ethylene oxide gas. If sterilization is not available, please conduct
immersion disinfection, rinse and dry them completely.
NOTE
Alcohol is flammable and must be used safely.
CAUTION
All accessories included are NOT immune to autoclave sterilization.